Effect of low tube voltage and low iodine concentration abdominal CT on image quality and radiation dose in children: preliminary study

  • 주제(키워드) Children , Computed tomography , Low tube voltage , Iodine concentration , Dose reduction
  • 주제(기타) Radiology, Nuclear Medicine & Medical Imaging
  • 설명문(일반) [You, Sun Kyoung] Chungnam Natl Univ, Dept Radiol, Coll Med, Daejeon, South Korea; [You, Sun Kyoung] Chungnam Natl Univ Hosp, Dept Radiol, Daejeon, South Korea; [Choi, Young Hun; Cheon, Jung-Eun; Kim, Woo Sun; Kim, In-One] Seoul Natl Univ, Dept Radiol, Childrens Hosp, Seoul, South Korea; [Choi, Young Hun; Cheon, Jung-Eun; Kim, Woo Sun; Kim, In-One] Seoul Natl Univ, Dept Radiol, Coll Med, Seoul, South Korea; [Cheon, Jung-Eun; Kim, Woo Sun; Kim, In-One] Seoul Natl Univ, Inst Radiat Med, Med Res Ctr, Seoul, South Korea; [Lee, So Mi] Kyungpook Natl Univ, Kyungpook Natl Univ Hosp, Sch Med, Dept Radiol, Daegu, South Korea; [Cho, Hyun-Hae] Ewha Womans Univ, Dept Radiol, Mokdong Hosp, Seoul, South Korea
  • 등재 SCIE, SCOPUS
  • 발행기관 SPRINGER
  • 발행년도 2019
  • 총서유형 Journal
  • URI http://www.dcollection.net/handler/ewha/000000171915
  • 본문언어 영어
  • Published As https://dx.doi.org/10.1007/s00261-019-01896-6
  • PubMed https://pubmed.ncbi.nlm.nih.gov/30683980

초록/요약

PurposeTo evaluate the image quality of a double-low protocol (low tube voltage and low iodine concentration) for abdominal CT in children.Materials and methodsThe double-low protocol was compared to the conventional protocol in pediatric patients weighing less than 40kg from May 2016 to December 2016. Double-low protocol (Group A, n=18): tube voltage, 70kVp; and iodine concentration,: 250mgI/mL versus Conventional protocol (Group B, n=13): tube voltage, 80-100kVp; and iodine concentration, 350mgI/mL. Mean attenuation, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were compared between the two groups. Image contrast, noise, beam-hardening artifacts, and overall image quality were subjectively scored. Reader performance for correctly differentiating two groups by visual assessment was evaluated. Radiation dose and total iodine load were recorded.ResultsThe mean attenuations of the portal vein and liver and the mean image noise in Group A were higher than in Group B (p=0.04, 0.03, 0.004, respectively). The mean SNR and CNR of the main portal vein and liver were lower in Group A without any statistically significant difference. There were no statistically significant differences between the two groups in qualitative analysis (image contrast, image noise, and overall image quality) with substantial agreement between the reviewers (weighted kappa values; 0.59-0.76). Significantly diminished radiation dose and iodine load were observed in Group A compared with Group B (25.0%, 36.8% reduction; p=0.007, 0.006, respectively).ConclusionThe double-low protocol was feasible for pediatric abdominal CT and reduced both radiation dose and iodine load, while maintaining image quality.

more